VWR CO2 Incubator Basic | 33
Start-up |
Chapter 5
CO
2
connection:
Figure 12 – Gas pressure hose installation
The gas supply from the gas supply system to the device is achieved using the supplied flexible
gas pressure hoses:
1. Connect gas pressure hose [1] to the sleeve of the gas supply system.
2. Remove protective cap [3] from sterile filter.
3. Slide hose clamp [2] onto gas pressure hose and connect hose to the sleeve [4] of the sterile
filter [5].
4. Secure gas pressure hose to the sleeve of the sterile filter using the hose clamp.
Power supply connection
WARNING
Electric shock:
Contact with current-carrying components may cause a lethal electric shock.
Before connecting the device to the power supply, check plug and connection line for damage.
Do not use damaged components when connecting the device to the power supply!
The device must be connected only to a correctly installed and grounded power supply source:
Fusing: T 16 A
Circuit breaker: G 16