Chapter 4
90 °C decontamination routine
The 90 °C decontamination routine is used to decontaminate the complete work space including
all installed components and sensors. During this routine, a moist and wet atmosphere with highly
decontaminating effect is created for 9 hours at a temperature of 90° C.
The effectiveness of the decontamination routine has been tested and certified by independent
institutes. Information about these tests is available at request from VWR.
The entire program run of the decontamination routine will take approx. 25 hours.
After the run has been completed, the device must be reactivated using the autostart routine.
NOTE!
Thermal protection:
If the thermal protection for the device responds, the 90 °C decontamination routine can only be
started after the cause of the failure has been repaired or reset.