VWR CO2 Incubator Basic | 27
Description of the device |
Chapter 4
Rear panel openings:
Figure 9 – Rear panel openings
A seal-able access port [1] allows cables, hoses or additional sensors to be routed into the work
space of the device. A pressure compensation opening with insert [2] at the rear panel of the
device allows a compensation between the pressures in the work space and in the operating room.
NOTE!
Operating conditions:
When accessories are to be operated in the work space of the CO
2
incubator, the ambient
condition requirements must be observed (see table below). The energy introduced into the work
space affects the beginning of the temperature control range. When additional heating sources are
introduced into the work space, condensation (e.g. at the glass door) may occur.
Introduced energy Control range of the temperature
General Example:
RT* = 21 °C
0 W RT + 3 °C 24 °C
5 W RT + 6.5 °C 27.5 °C
10 W RT + 9.5 °C 30.5 °C
15 W RT + 13 °C 34 °C
20 W RT + 16 °C 37 °C
*RT = Ambient temperature