| Maintenance
64 | CO2 Incubator Basic VWR
Chapter 10
Measurement example:
Temperature nominal value: 37 °C
Reference temperature: 36.4 °C
NOTE!
Excessive work space temperature:
Excessive work space temperature after the calibration can be reduced by leaving the doors open
for approx. 30 seconds!
Temperature calibration procedure
1. Activate calibration:
Keep the key depressed for 5 seconds.
All operating panel indicators flash.
2. Indicate the nominal value:
Press the key.
The preset value of 37 °C is displayed.
3. Enter the measured value (destination value):
Press the + keys.
or
Press the + keys.
Destination value e.g. 36.4 °C
4. Accept the destination value:
Press the key.
The temperature display momentarily shows “CAL”,
Then the corrected actual value (measured destination value 36.4 °C) is displayed.