VWR CO2 Incubator Basic | 63
Maintenance |
Chapter 10
Comparison measurement procedure:
Figure 19 – Temperature calibration
1. Turn device on using power switch.
2. Set temperature nominal value and allow device to be heated. This may take up to several
hours.
3. Place measuring instrument [3] onto the A center area of the work space.
Alternatively, a temperature sensor may be positioned in this location. Route the connecting
cable either through the measurement opening [2] in the glass door or through the access port
[1] at the rear panel of the device.
4. Close doors.
5. Wait until the temperature value displayed on the measuring instrument has stabilized.
6. Calibrate temperature control as described in Section “Temperature calibration procedure”.