What to do if there is no clean water in the tank of my W&H Laboratory Equipment, but the automatic water filling does not fill the water?
- KKelly BullockSep 12, 2025
If there is no clean water in the tank and the automatic water filling doesn't work, it could be due to: * Water fill system not connected: Connect the water fill system to the sterilizer. * Water was temporarily unavailable: Switch the sterilizer OFF and then ON again. Check the external water supply system and check for water leaks from the sterilizer. * Faulty MIN water level sensor in the clean water tank: Call service.