Maintenance
Lexa PLUS| Instructions for Use |RIS-311 ENG Rev03 |29/08/2022 |© W&H Sterilization Srl 73
1200-cycle or yearly maintenance
REPLACING THE DOOR GASKET
Notice: The door gasket needs to be replaced every 1200 sterilization cycles or once a year, whichever come first. A replacement message alerts
when replacement is due. If the consumables is replaced prior to the message, you have to reset the consumable cycle counter.
Proceed as follows:
1 Open the chamber door.
2 Remove the used door gasket by hand.
3 Carefully clean the seal seat and the inside face of the chamber door.