98 Lexa PLUS| Instructions for Use |RIS-311 ENG Rev03 |29/08/2022 |© W&H Sterilization Srl
Technical data
DISTILLED OR DEMINERALIZED WATER
Water quality
See the Test instructions service book (conductivity : < 10μS/cm,
Total Dissolved Solids: < 6.5 ppm)
Average water
consumption
0.07 to 0.23 gal/cycle (0.26 to 0.85 litres/cycle)
Tank volume
Clean water 1.27 gal (4.8 l)
Min. water charge 0.22 gal (0.82 l)
Used water 1.35 gal (5.11 l)
*: usable space with standard rack and trays. With optional racks and trays,
see Accessories, spare parts, consumables.
Recommendations for validation
TEST VALIDATION POINTS
Part Description
Hottest points
Coldest points