70 Lexa PLUS| Instructions for Use |RIS-311 ENG Rev03 |29/08/2022 |© W&H Sterilization Srl
800-cycle or biannual maintenance
SEQUENCE OF PROCEDURES TO CLEAN THE WATER TANK
To clean the water tank, proceed as follows:
1 "Prepare the sterilizer for cleaning the water tank" below "Prepare the sterilizer for cleaning the water tank" below.
2 "Access to the water tank" on the next page" "Access to the water tank" on the next page.
3 "Clean the water tank" on page72
PREPARE THE STERILIZER FOR CLEANING THE WATER TANK
Proceed as follows:
1 Switch OFF the sterilizer and disconnect the mains cable.
2 Completely drain both water tanks (see "Draining the used and clean water tank"
on page75).
3 To drain the detergent solution during the subsequent cleaning, leave the drain
tube attached to the drain connection of the tank you want to clean.