Maintenance
72 Lexa PLUS| Instructions for Use |RIS-311 ENG Rev03 |29/08/2022 |© W&H Sterilization Srl
CLEAN THE WATER TANK
Notice: Do not touch the water level sensors. If misplaced or misaligned from their original position, the operation of the sterilizer could be
impaired.
Proceed as follows:
1 Clean the internal tank surfaces with a soft sponge moistened with water. For
better cleaning results, clean with W&H MC-1000 cleaning solution. Use a small
non-abrasive brush to clean the areas that are difficult to reach.
Notice: never use any other disinfectant, detergent or abrasive product, as they
might result aggressive for the tank material.
2 Rinse the internal tank surfaces thoroughly, until all residuals of dirt and
detergent have been eliminated.
3 Dry the internal tank surfaces.
5 Remove the internal filter.
6 Clean the metal cartridge of the internal filter with tap water.
7 Reposition the internal filter.
8 Detach the drain tube.