Maintenance
64 Lexa PLUS| Instructions for Use |RIS-311 ENG Rev03 |29/08/2022 |© W&H Sterilization Srl
CONTENTS
This section deals with the following subjects:
Warnings for maintenance operations 64
User maintenance 64
Monthly or 50-cycle maintenance 66
400-cycle maintenance 69
800-cycle or biannual maintenance 70
1200-cycle or yearly maintenance 73
Extraordinary maintenance 75
Disposal 76
Warnings for maintenance operations
WARNINGS
WARNING! Turn the sterilizer OFF and remove the power
cord before beginning any maintenance. Follow all health,
safety, cross-infection and cross-contamination
protocols.
Maintenance operation shall be done at illumination level
of 215 lx (±15 lx) to 1500 lx (±15 lx).
Before making any operation, ward off unauthorized
personnel from the working area.
CAUTION! Before accessing the chamber and the connected
parts, be sure that the sterilizer is cold.
Notice: Follow the instructions in this chapter when carrying out any
maintenance on the sterilizer.
User maintenance
MAINTENANCE BY THE USER
Frequency
1
Cycles
1
Operation
Monthly 50
Cleaning the door gasket and the chamber face
side. See "Cleaning the door gasket and the
chamber face side" on page66
Clean the chamber, trays and the rack. See
"Cleaning the chamber and the chamber
accessories" on page67
Cleaning the chamber filters. See "Cleaning the
chamber filters" on page68
Cleaning the external surfaces of the sterilizer.
See "Cleaning the external surfaces of the
sterilizer" on page68
6 month 800 Clean the clean water tank. See "800-cycle or
biannual maintenance" on page70