Maintenance
68 Lexa PLUS| Instructions for Use |RIS-311 ENG Rev03 |29/08/2022 |© W&H Sterilization Srl
CLEANING THE CHAMBER FILTERS
Proceed as follows:
1 Allow the sterilization chamber to cool down.
2 Empty the sterilizer chamber by removing the trays and the rack.
3 Remove the filters at the bottom of the chamber (bottom/center) by unscrewing
them.
4 Rinse the filters with tap water.
5 Insert the filters in their original position, screwing them into place.
CLEANING THE EXTERNAL SURFACES OF THE STERILIZER
Proceed as follows:
1 Clean all external sterilizer covers with a slightly damp cloth moistened with
water. For better cleaning results, clean with W&H MC-1000 cleaning solution.
Notice: never use any other disinfectant, detergent or abrasive product, as they
might result aggressive for the external covers and damage them.