EasyManua.ls Logo

Waters ACQUITY RDa - Page 54

Waters ACQUITY RDa
107 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Figure 3–19: Fitting the probe capillary assembly
PEEK fitting Probe capillary
assembly
Probe fitting Capillary
8. Fit the PEEK fitting end of the probe capillary assembly to the port S on the divert valve
and secure until finger-tight.
Important: To avoid dead volumes, ensure that the capillary bottoms out in the valve.
9. To avoid trapping and damaging the capillary in the instrument door, slide the capillary
beneath the retaining clips.
10. Close the instrument door.
11. In the UNIFI software, click Operate to put the instrument into Operate mode.
August 20, 2021, 715005003 Ver. 04 (Previously released as Revision C)
Page 54

Table of Contents

Related product manuals