16 Assembly
EN
2. Screw the pressure reducer with the corrugated
union nut to the cylinder valve. Tighten the
union nut hand-tight.
3. Screw the pressure tube with the G 3/8 union
nut to the pressure reducer outlet.
4. Screw the other end of the pressure tube to
oxygen inlet
1 of the module.
4.3 External oxygen connection
In addition to oxygen inlet 1, you can use external
oxygen connection
8 to tap off or feed in oxygen.
Tapping off oxygen
You may not operate a module at the same time as
tapping off oxygen, otherwise the specified
technical data will not be maintained.
Feeding in oxygen
You can use the external oxygen feed for example to
allow continued working with the modules when
the oxygen cylinder is being changed.
Due to the nonreturn valves in oxygen inlet
1,
oxygen should always be fed in at the module fitted
on the extreme left.