Suction MODULE 25
EN
Method
1. Switch on the vacuum at toggle switch 4.
2. Close the fingertip regulator and hold the
opening of the end-piece closed so that no more
air can flow through it.
3. Regulate the vacuum using dial
5. There are
three revolutions of the dial between minimum
and maximum.
4. Check the value set at pressure gauge
2.
5. Now you can aspirate.
Keep checking the filling level of the oxygen
cylinder from time to time.
6.3 Stop aspiration
1. Check the oxygen reserve at the contents pressure
gauge of the pressure reducer.
2. Close the oxygen supply valve and wait until the
system is depressurized.
3. Turn dial
5 to the left-hand stop (to “zero”). This
prevents the pressure gauge being damaged by
hammering when the device is switched back
on.
4. Switch the module off using toggle switch
4.
5. Clean and disinfect/sterilize the device and the
device parts (see “11. Hygiene treatment” on
page 39).
6. Then perform a function check (see “12.
Function check” on page 42).
Medium
Vac
2,7 - 6 bar
O
2
l/min
I
0
MODUL
Suction