40 Hygiene treatment
EN
11.2 Cleaning, disinfecting and sterilizing
The individual parts can be cleaned, disinfected and
sterilized as listed in the table below.
Follow the instructions for use for the disinfectant
used. We recommend TERRALIN. It is recommended
that suitable gloves (e.g. household or disposable
gloves) are used for disinfecting.
Parts Clean Disinfect
Washing
machine cycle
Sterilize
Modules
Wipe down with a
damp cloth
Disinfect by
wiping
Not permitted
Oxygen fittings
Clean cloth,
possibly
moistened with
clean water
Not permitted
OMNIVAC secretion aspiration system
Lid of aspirate
bottles
In hot water with
a mild household
detergent
Immerse in
dilute solution
(1)
Not permittedBall of overflow
cutout
Connectors
Aspirate bottles
Washing cycle
up to 95 °C
Hot-steam
sterilization
up to 134 °C
(2)
Aspirating tube
Intermediate
tube
Vacuum tube
End-piece with
fingertip
regulator
Disposable items, re-use not permitted, replace for every patient
Bacteria filter
Cannula/mask for
inhalation, disposable