32 CPAP MODULE
EN
9. CPAP MODULE
9.1 CPAP function
You can use the CPAP MODULE to perform CPAP
therapy in conjunction with a CPAP valve approved
by WEINMANN Emergency.
Conditions of use
Therapy may only be carried out on emergency
patients who are breathing spontaneously.
In the event of respiratory failure, CPAP therapy is by
no means adequate and artificial ventilation will be
required here.
The patient is supplied with oxygen via a CPAP valve
and a breathing mask.
Risk to patient!
Ensure that the free end of the CPAP valve is not
covered or sealed off.
The patient must be able to breathe freely.
9.2 Operation
CPAP therapy
1. Slowly open the oxygen supply valve.
The contents pressure gauge indicates cylinder
pressure.
2. Calculate remaining operating time (see
“Calculating filling level” on page 43).