Technical data 57
16. Technical data
* If ambient conditions and feed pressures differ from these, the
specified performance data may also deviate.
The right to make design modifications is reserved.
Oxygen
MODULE
Suction
MODULE
Combi
MODULE
Interface
MODULE
CPAP
MODULE
Dimensions
W x H x D in mm
100 x 130 x 90 100 x 172 x 90 100 x 145 x 90
Weight 0.9 kg 0.9 kg 0.95 kg 0.65 kg 0.9 kg
Temperature range
– operation
– storage
–18 °C to +60 °C
–40 °C to +70 °C
Operating gas Medical oxygen
Operating pressure 2.7 – 6 bar possible, recommended: 4.5 bar
Req. gas quantity ≥ 80 l/min O
2
Compressed gas
connection
External thread G3/8; NIST adapter available on request
Oxygen output
Infinitely
adjustable
from 0 to
15 l/min
Infinitely
adjustable
from 0 to
15 l/min
Infinitely
adjustable from
0 to approx.
43 l/min
Tolerance range
At 0–3.9 l/min
±15 %
At 4–15 l/min
±10 %
At 0–3.9 l/min
±15 %
At 4–15 l/min
±10 %
Maximum CPAP
(at feed pressure
4.5 bar, ambient
conditions 1013 hPa/
21 °C )
Approx.
15 ±3 hPa*
Class to 93/42/EEC IIa
Classification to
EN ISO 10079-3
Medium vacuum
Accuracy of pressure
display
Class 1.6
Flow outlet 9/16-UNF 9/16-UNF DIN coupling 9/16-UNF
Standards applied EN 1789