EasyManua.ls Logo

Weinmann Module System - Assembling the OMNIVAC Secretion Aspiration System

Weinmann Module System
68 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Hygiene treatment 41
EN
(1) After disinfecting the parts, rinse them thoroughly in
distilled water and then leave them to dry.
(2) Hot steam sterilization at 134 °C in devices to EN 285,
dwell time 5 minutes.
11.3 Assembling the OMNIVAC
secretion aspiration system
Following the hygiene treatment, reassemble the
parts in reverse sequence (see “11.1 Preparing the
OMNIVAC secretion aspiration system” on
page 39).
Mask for inhalation,
reusable
In hot water with
a mild household
detergent
Immerse in
dilute solution
(1)
Washing cycle
up to 95 °C
Hot-steam
sterilization
up to 134 °C
(2)
CPAP valve Disposable item, re-use not permitted, replace for every patient
Parts Clean Disinfect
Washing
machine cycle
Sterilize

Table of Contents

Related product manuals