4 Function check
2. Open zippers fully on both sides.
Please note: The zippers must not stick.
3. If the emergency backpack does not work perfectly: Replace
emergency backpack.
4.4 Checking charging interface
1. Power the portable unit via the charging interface (see 5.4.2,
p. 21).
Please note:
• The LED for the external power supply on the fitted device
must illuminate.
• The battery status display must illuminate or flash.
2. Check that the charging interface port is secured firmly.
Please note:
• The charging interface port must not move.
• The charging interface port must fit snugly against the
emergency backpack.
3. If the charging interface does not function perfectly: Have
WEINMANN Emergency repair or replace the charging
interface.
4.5 Checking lock on LIFE-BASE 1NG XS
1. Insert portable unit into holding bracket.
Please note:
• The portable unit must sit firmly in the holding bracket.
• The red control pin of the lock cross-bar must be flush with
the top edge of the lock housing (see 5.4.2, p. 21).
2. Hold with one hand through the handle.