WM 66531 2023-01
RESCUE-PACK Plus EN 41
9 Faults
9Faults
If you are not able to remedy faults at once with the aid of the
table, you should contact the manufacturer
WEINMANN Emergency or an WEINMANN Emergency authorized
dealer to have the emergency backpack repaired. To avoid more
serious harm, do not continue using the emergency backpack.
Fault Cause Remedy
Device cannot be secured or is not
seated in the intended position
Device not inserted properly in
holding bracket
Check device is seated properly.
Holding bracket not secured
Check holding bracket is seated
properly.
Holding bracket defective Replace holding bracket.
Latch for holding bracket defective Replace latch for holding bracket.
Mounting loops for holding
bracket in the emergency
backpack defective
Replace emergency backpack.
Portable unit defective
Check portable unit for damage
and replace if necessary (see 4,
p. 16). You should also follow the
instructions for use of the portable
units.
Device not charging
Crimp connector not connected
correctly
Connect crimp connector correctly
(see 8.3, p. 37).
Charging interface port
contaminated
Clean charging interface port (see
7, p. 30).
Charging interface port defective
Replace charging interface port
(see 6.5, p. 28), (see 8.3, p. 37).
Connection line for charging
interface defective
Replace connection line for
charging interface.
Charging interface not connected
to device
Connect charging interface to
device.
Charging adapter not connected
correctly
Connect charging adapter
correctly (see 5.3, p. 20).
Charging adapter contaminated
Clean charging adapter (see 7,
p. 30).
Charging adapter defective Replace charging adapter.