WM 66531 2023-01
RESCUE-PACK Plus EN 35
7 Hygienic reprocessing
7.7 Cleaning and disinfection plan
Carry out hygienic reprocessing as per the table below. Please note
the intended intervals (see “7.1 Intervals”, page 30).
The applicable instructions are those in the instructions for use
enclosed by the manufacturers of the individual parts. Follow these
instructions for use.
Part
Manual
cleaning
Disinfection
by wiping
Disinfection
by immersion
Mechanical
reprocessing
Sterilization
Emergency
backpack
Wipe down
smooth surfaces
with neodisher
®
MediClean forte
(Dr. Weigert).
Dose: 10 ml/l.
Treat all surfaces
at least twice
until they look
clean.
Not permitted Not permitted
Prewash
5minutes at
20 °C,
main wash
30 minutes at
30 °C.
Cleaning agent:
Turbo Usona
(Ecolab),
Disinfectant:
Turbo Oxysan.
Not permitted
Accessories
bags
Holding bracket
Wipe down with
neodisher
®
MediClean forte
(Dr. Weigert).
Dose: 10 ml/l.
Treat all surfaces
at least twice
until they look
clean.
Wipe down all
surfaces once
with Incidin
™
OxyWipe S (Eco-
lab).
Exposure time:
5minutes.
Not permitted Not permitted Not permitted
Charging
interface
Wipe down with
neodisher
®
MediClean forte
(Dr. Weigert).
Dose: 10 ml/l.
Treat all surfaces
at least twice
until they look
clean.
Wipe down with
Incidin
™
OxyWipe S (Eco-
lab).
Exposure time: 5
minutes.
Treat all surfaces
at least twice
until they look
clean.
Not permitted Not permitted Not permitted
Cables
Portable unit
Follow the instructions for use of the portable unit and the device.
Device