7 Hygienic reprocessing
4. For the LIFE-BASE light XS holding bracket: Remove retaining
strap with magnetic fastener from holding bracket.
7.3 Cleaning parts manually
Requirement • The parts exhibit visible dirt.
• Hygienic reprocessing is prepared (see 7.2, p. 30).
1. Refer to the cleaning and disinfection plan for details of
approved parts, approved agents, and the dose (see 7.7,
p. 35).
2. Prepare the cleaning solution as specified by the cleaning agent
manufacturer.
3. To remove all visible dirt: Brush parts thoroughly using a
commercially available soft brush suitable for plastic and
wetted with the cleaning agent.
Please note: Uneven surfaces and grooves (e.g., hook and loop
fasteners, zippers, buckles) must be brushed especially
thoroughly.
4. Wipe down parts using a clean, lint-free disposable cloth
moistened with cleaning solution.
Please note:
• Use a new disposable cloth for every cleaning process.
• All surfaces must be wiped carefully.
• All surfaces must be wetted with cleaning solution.
• The cleaning cycles specified in the cleaning and
disinfection plan must be observed.
• Uneven surfaces and grooves need wiping over thoroughly.
5. For cables: Wrap disposable cloth around cable and pull the
cable through the disposable cloth.
6. In the event of persistent visible soiling: Repeat manual
cleaning.
7. Wipe down parts with a damp cloth to remove residues of the
cleaning agent.