39 - EMC Guidelines and Manufacturer Declaration
Table 2 – Guidelines and Manufacturer Declaration
Electromagnetic immunity – for all ME devices and ME systems
Guidelines and Manufacturer Declaration
-
Electromagnetic immunity
The ABPM 7100 is intended for operation in an electromagnetic environment as specified below. The customer or
user of the ABPM 7100 should ensure it is used in such an environment.
Immunity tests Test level Compliance levels
Electromagnetic Environment -
Guidelines
Electrostatic discharge
(SD)
according to IEC
61000-4-2
± 8 kV
Contact discharge
± 15 kV
Air discharge
± 8 kV
Contact discharge
± 15 kV
Air discharge
Floors should consist of wood or cement
or ceramic tiles. If the floor consists of
synthetic materials, relative humidity must
be at least 30%.
Rapid transient
electrical
disturbance/bursts
according to IEC
61000-4-4
± 1 kV
100 kHz repetition
rate
± 1 kV
100 kHz repetition
rate
Surges
according to IEC
61000-4-5
± 1 kV
Line-to-line voltage
± 2 kV
Line-to-earth voltage
Not applicable
Not applicable
The ABPM 7100 does not have an AC
power supply.
Magnetic field in
supply frequency
(50/60 Hz)
according to IEC
61000-4-8
30 A/m 30 A/m Magnet fields in network frequency should
correspond with the typical values found
in business and hospital environments.
Voltage drops,
short interruptions
and fluctuations in
supply voltage
according to IEC
61000-4-11
0% UT
for 0.5 cycles
Not applicable The ABPM 7100 does not have an AC
power supply.
0% UT
for 1 cycle
Not applicable
70% UT
for 25/30 cycles
Not applicable
0% UT
for 250/300 cycles
Not applicable
NOTE UT is the AC voltage before the application of the test levels.
EMC Guidelines and Manufacturer Declaration - 40
Table 3 - Electromagnetic immunity for casings designed to shield against high-frequency wireless
communication devices
Guidelines and manufacturer declaration – electromagnetic immunity
The ABPM 7100 is intended to be operated in the electromagnetic environment specified below. The customer
or the ABPM 7100 user should ensure that it is used in such an environment.
Emitted interference
measurement
Test level Compliance level
HF radiated disturbances in
accordance with IEC 61000-4-3
380 - 390 MHz
27 V/m; PM 50%; 18 Hz
430 - 470 MHz
28 V/m; (FM ±5 kHz, 1 kHz sine)
PM; 18 Hz
704 - 787 MHz
9 V/m; PM 50%; 217 Hz
800 - 960 MHz
28 V/m; PM 50%; 18 Hz
1700 - 1990 MHz
28 V/m; PM 50%; 217 Hz
2400 - 2570 MHz
28 V/m; PM 50%; 217 Hz
5100 - 5800 MHz
9 V/m; PM 50%; 217 Hz
380 - 390 MHz
27 V/m; PM 50%; 18 Hz
430 - 470 MHz
28 V/m; (FM ±5 kHz, 1 kHz sine)
PM; 18 Hz
704 - 787 MHz
9 V/m; PM 50%; 217 Hz
800 - 960 MHz
28 V/m; PM 50%; 18 Hz
1700 - 1990 MHz
28 V/m; PM 50%; 217 Hz
2400 - 2570 MHz
28 V/m; PM 50%; 217 Hz
5100 - 5800 MHz
9 V/m; PM 50%; 217 Hz
Table 4 – Guidelines and Manufacturer Declaration
Electromagnetic immunity for ME devices or ME systems that are not life-supporting
Guidelines and Manufacturer Declaration
-
Electromagnetic immunity
The ABPM 7100 is intended for operation in an electromagnetic environment as specified below. The customer or
user of the ABPM 7100 should ensure it is used in such an environment.
Immunity tests
Test
level
Compliance level
Radiated disturbance variables
according to
IEC 61000-4-3
10 V/m
80 MHz to
2.7 GHz
10 V/m
Conducted disturbance variables
according to
IEC 61000-4-6
Not applicable