bags and you are not sure whether they are static-sensitive, assume that they are static-
sensitive and handle them accordingly.
Perform all service procedures in a static-protected environment. Always use
techniques and equipment designed to protect personnel and equipment from
electrostatic discharge.
Remove static-sensitive components and assemblies from their static-shielding bags
only at static-safe workstations—a properly grounded table and grounded floor mat—
and only when you are wearing a grounded wrist strap (with a resistor of at least 1
megohm in series) or other grounding device.
Use only grounded tools when inserting, adjusting, or removing static-sensitive
components and assemblies.
Remove or insert static-sensitive components and assemblies only with device
power turned off.
Insert and seal static-sensitive components and assemblies into their original static-
shielding bags before removing them from static-protected areas.
Always test your ground strap, bench mat, conductive work surface, and ground
cord before removing components and assemblies from their protective bags and
before beginning any disassembly or assembly procedures.
6 Safety CP 150 12-lead resting electrocardiograph

Table of Contents

Other manuals for Welch Allyn CP 150

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Welch Allyn CP 150 and is the answer not in the manual?

Welch Allyn CP 150 Specifications

General IconGeneral
TypeElectrocardiograph
BatteryRechargeable lithium-ion
ConnectivityUSB, Ethernet
Paper Size8.5 x 11 inch
ECG Channels12
ECG InterpretationOptional
Display7-inch color touchscreen

Related product manuals