Upgrade your existing CP 150 device software version
through the Welch Allyn Service Tool
Note A USB cable is required to perform the software upgrade.
Note Connect the CP 150 to an AC power outlet before upgrading the software.
1. Download the Welch Allyn Service Tool and Partner Connect and install on your PC.
2. Follow the instructions to set up a user name and password.
3. Plug your CP 150 device into a USB port on your PC running the Welch Allyn Service
Tool and turn on the device. Plug your CP 150 device into AC power.
4. If the Welch Allyn Service Tool is not still open, go to the Windows Start menu, and
then select All Programs > Welch Allyn > Welch Allyn Service Tool.
5. Log in to the Welch Allyn Service Tool.
6. Highlight the CP 150 to choose it from the Device list.
7. Click Select.
8. Click the Upgrade tab and then click Check for new upgrades.