Perform the full-functional tests
Complete the full-functional CP 150 tests to verify device functionality.
Note The functional verification tests help to confirm the proper operation of the
device and its options. The tests may also be useful as a diagnostic tool to
help isolate a malfunction.
The following off-the-shelf tools are required to complete the functional verification tests:
Standard ECG Patient Simulator (example: Model 430B, Medi Cal Instrument, Inc. or
equivalent)
USB Cable (Type A connector to mini-B connector)
Standard USB Thumb drive (256M to 64Gb, recommend 2Gb)
RJ45 Ethernet cable (less than 3 M in length)
1. From the Home screen, touch the Settings tab.
2. Touch the Advanced tab.
3. Enter 6345 as the access code and touch OK.
The General tab appears at the bottom of the screen and the Regional tab appears at
the top of the screen.
4. Touch the Service tab.
5. Touch the Self-tests tab.
Test the hardware components.
6. Touch CP 150 to perform the 9 hardware tests.
50 Service CP 150 12-lead resting electrocardiograph