Steam Sterilizer Operation Manual
TANDA Model REV-E Page of 43
9
Only an approved technician using original spare parts can carry out repairs and
maintenance.
In case of transportation, drain both water tanks completely, allow the sterilization
chamber to cool down and preferably use the original packaging.
The articles under sterilization should be removed by special tools when the temperature
over 40.
Picking-up the sterilizing trays should use the special tools provided.
During the transportation, the sterilizer should be carried by two people in case of turning
over.
Notice! This product can’t be put on the place that is not easy to cut off power supply.
Prohibit covering the lid of water tank during usage.
2.3 Safety Parts
Temperature Protection
Part Name
Function
Temperature Protector
(Steam Generator)
Cut off current when the steam generator temperature is too high.
Temperature Protector
(Heating Ring)
Cut off current when the heating ring temperature is too high.
Electricity Protection
Part Name
Function
Double Fuse
Cut off current when the connected power is too high or unstable.
Electronic Filter
Filter the electromagnetic interference during working.
Mechanical Protection
Part Name
Function
Jiggle Switch
Ensure the door closed completely, avoiding safety risk.
Tray Tong
Avoid scald when removing articles from the chamber