ATTENTION: Make sure that use distilled water properly in order to extend the sterilizer
life-time.
Do not rave about the sterilizer when the tank has been filled.
8.5 Other Maintenance
Changing Door Gasket
Tool: A plain screw driver without sharp head is needed.
Cut off the power supply, ensure that the sterilizer is cool and depressurized.
1. Hold verge of the seal by one hand softly, and another hand should be inserted the
screwdriver into the gap between gasket and door, take out the seal slowly.
2. Once you take out one part of the seal, you can draw out the whole seal slowly. After
taking out the seal, please check and clean the groove of gasket, so does the gasket,
please replace it if there is some damage.
3. Fix the clean gasket in initial door groove. At first, embed 4 spots equably into groove
when fix the gasket, then embed the other parts. Next, embed it equably by hands.
4. The inner edge of gasket may be ectropion during embedding it in the door groove. You’d
better to tight it back to the groove by using screwdriver carefully.