Steam Sterilizer Operation Manual
TANDA Model REV-E Page of 43
24
6.3 Alarming if Used Water Reservoir is Full
The OUTindicator light illuminates during the cycle, which means that the used water tank
needs to be drained.
Connect the drainage tube to used water outlet, and used water will be drained automatically.
Generally, the maximum temperature of drained water should be under 70. If it is higher, you
need to check whether the fan work normally, or contact the local distributor immediately.
6.4 Selecting Sterilizing Program
Select the required sterilization program which you need. When you choose, the
corresponding indicator light will be illuminated.
6.5 Loading Articles
Articles should be put on the trays with some gap between each articles so that the steam can
be ventilated freely. Please use the attached removal tool to load trays into chamber to avoid
scalding. (pic 6-4)
pic 6-4
Tray arrangement note:
Read the following instructions for proper placing of articles and material.
Make sure that the articles of different materials are separated and placed on different
trays.