8.6 Servicing by the Approved Technician
Service is essential for consistently effective sterilization.
We recommend servicing by an approved technician every 2 years.
Check-list:
1 Checking the solenoid valves.
2 Checking the water pump.
3 Checking the vacuum pump.
4 Checking the distilled water drain valve and the used water drain valve.
5 Checking the relief valve.
6 Checking the door locking system.
7 Checking the probe of the pressure and temperature.
8 Checking the probe of the water in the sterilization chamber.
9 Checking the electrical connections.
10 Checking the hydraulic connections.
11 Checking the safety thermostat.
12 Cleaning the sterilization chamber.
13 Cleaning the trays and the tray rack.
14 Cleaning the reservoirs.
15 Replacing the water filter.
16 Replacing the air filter.
17 Replacing the door gasket.