Steam Sterilizer Operation Manual
TANDA Model REV-E Page of 43
18
4.7 Sterilization Cycle
-0.8
-1.0
1
2
3 4 5
6 7
kPA
Time
0.8
2.0
P
r
e
s
s
u
r
e
END
1-7 Entire duration
1-2 Pre-heating
2-3 Pre-vacuum
3-4 Sterilizing
4-5 Air-discharging
5-6 Drying
6-7 Stabilizing
Table—Types of sterilization cycles
Type
Purpose description
B
The sterilization of all wrapped or non-wrapped, solid, hollow load products type
A and porous products as represented by the test loads in this standard.
S
The sterilization of products as specified by the manufacturer of the sterilizer
including non-wrapped solid products and at least one of the following: porous
products, small porous items, hollow load products type A, hollow load products
B, single wrapped products, multiple-layer wrapped products.
NOTE 1
The description identifies ranges of products and test loads.
NOTE 2
Non wrapped sterilized instruments are intended either for immediate use or for
non sterile storage, transport and application (e.g. to prevent cross infection).