Steam Sterilizer Operation Manual
TANDA Model REV-E Page of 43
17
Maximum capacity of one tray: 1000g
The frequency of water draining: once a day, drain the water once you find “waster water over”
during operation.
The maximum duration of using loading test: 90mins
The maximum thermal radiation energy under the condition of 20~26: <2000J
Sterilizer Chamber
Material: stainless steel (for medical)
Max. work pressure: 2.5bar
Min. work pressure: -0.9bar
Max. temperature: 145
Chamber volume: 18L(Φ245×352mm) 23L(Φ245×450mm)
Loading size: 18L (198×204×285mm) 23L(198×204×385mm)
Max. loading weight: 18L (3.07kg/cm
2
) 23L (3.21kg/cm
2
)
Working pressure/temperature: 1.10~1.30bar/121~122; 2.10~2.30bar/134~135
Water volume for one cycle: 0.16L~ 0.18L
Sterilizer Steam Safety Valve
Safety release pressure: 2.45bar
Max. working temperature: 160
Water Tank
Main water tank volume: 18L (3.5L) 23L (4L)
CAUTION: Water added into main water tank must be distilled water!
Water temperature must be under 40.
Test Method
Vacuum Test
B&D Test
Helix Test