Steam Sterilizer Operation Manual
TANDA Model REV-E Page of 43
Appendix 4 The Standards of Testing
The exterior of sterilizer should be tidy and mustn’t have
disfigurements, e.g.deflection, hollowness, collision, nick, sharp edge.
The cover plate should be assured to disassemble easily in order to
repair the equipment.
The digit and letter in screen should be legible.
The electroplate should accord with YYOO76-1992 class 2, which for
the request of aspect.
The printer components should accord with YY1055-1999 the class Ⅱ
which for the request of aspect.
On the normal condition, if the sterilizer door hasn’t been locked
tightly, the program can not start.
The sterilizer should ensure that the door can’t be opened when
chamber pressure is greater than 0.27kPa.
The sterilizer must install a safety valve, safety valve opening
pressure 0.27kPa ± 0.01kPa, and automatically open when reaching
the set value.
Sterilizer should have the pre-established program about 121℃ and
134℃, dressing and instruments.
The control system in sterilizer should limit the steam which in the
chamber be controlled at the highest average temperature in ±3℃ of
pre-establish station. And ensure the temperature value accords with
the pressure controlling value.
Able to timing control of the sterilization and drying, and the error
should not be greater than 10% of the preset value.
Buttons and switches should be flexible and reliable on the sterilizer.
The indicators and displays of sterilizer should show the states of
every sterilizing procedure exactly. Under the normal situation,
sterilizer should indicate:
a) Chamber temperature
b) Chamber pressure
c) Sterilizer working state
d) Water level state
e) State of door
In the condition of the vacuum - 0.07Mpa, the sterilizer shouldn’t leak
0.13kPa/min.
The sterilizer can’t leak under the work pressure