SCPI Commands for Digital Interfaces
TM-XDOP-01XN 4–27
Program V,I,P
SCPI
To change setpoints:
SOUR:VOLT <voltage>
SOUR:CURR <current>
SOUR:POW <power>
To check setpoints:
SOUR:VOLT?
SOUR:CURR?
SOUR:POW?
To set a triggered setpoint:
SOUR:VOLT:TRIG <voltage>
SOUR:CURR:TRIG <current>
SOUR:POW:TRIG <power>
See “Triggering Commands” on page 4–36 for more information.
To check a triggered setpoint:
SOUR:VOLT:TRIG?
SOUR:CURR:TRIG?
SOUR:POW:TRIG?
To set limits:
SOUR:VOLT:LIM:HIGH <voltage>
SOUR:VOLT:LIM:LOW <voltage>
SOUR:CURR:LIM:HIGH <current>
SOUR:CURR:LIM:LOW <current>
SOUR:POW:LIM:HIGH <power>
SOUR:POW:LIM:LOW <power>
If the high end of the range was set to 5 volts, the command, SOUR:VOLT 10
would return an error.