Multi 3410 What to do if...
110
ba75863y18 01/2015
Data transmission to
USB memory device
does not work
Data transmission to
USB printer device does
not work
Error message,
Only use rechargeable
batteries!
Ni-MH 1.2 V, >2100 mAh
CAUTION
Unsuitable batterie types can cause damage during the charging
process.
Cause Remedy
Connected USB memory device
has not been recognized, e. g.
the USB memory device has
been formattet to a file sys-
tem which is not supported
(e. g. NTFS)
Use other USB memory
device
Reformat the USB memory
device to FAT 16 or FAT 32
file system
(Caution
: Reformatting
erases all data on the USB
memory device. Back up all
data before reformatting.)
Cause Remedy
The USB-B interface is con-
nected to a PC
Disconnect the PC from the
USB-B interface
Connected USB memory device
does not work
Use a suitable USB printer
device (see section 4.12.3)
Check printer settings (see
section 4.12.3)
No power pack connected Connect the power pack
Cause Remedy
A voltage that is unusual for
NiMH rechargeable batteries
was found.
Disconnect the power pack
from the meter.
Check whether there are
suitable rechargeable batter-
ies (Ni-MH 1.2 V, >2100
mAh) in the battery compart-
ment.
Unsuitable batteries:
Insert suitable recharge-
able batteries and recon-
nect the power pack.
Suitable batteries:
Connect the power pack
and close the error mes-
sage with <[ESC]> or <M>.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Xylem wtw Multi 3410 and is the answer not in the manual?

Xylem wtw Multi 3410 Specifications

General IconGeneral
BrandXylem
Modelwtw Multi 3410
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals