Multi 3410 Operation
102
ba75863y18 01/2015
4.12.6 MultiLab Importer
With the aid of the MultiLab Importer software, you can record and eval-
uate measurement data with a PC.
Note
More detailed information can be found in the MultiLab Importer oper-
ating manual.
4.13 MultiLab User
The user administration function is activated via the MultiLab User PC
software.
The MultiLab User PC software can be found on the enclosed CD-
ROM.
Note
More detailed information can be found in the MultiLab User operating
manual.
4.14 Reset
You can reset (initialize) all sensor settings and sensor-independent
settings separately from each other.
4.14.1 Resetting the measurement settings
Note
The calibration data are reset to the default settings together with the
measuring parameters. Recalibrate after performing a reset.
pH The following settings for pH measurements are reset to the default
settings with the Reset function:
Setting Default settings
Buffer AutoCalTEC
Cal. interval 7 d
Unit for slope mV/pH
Measured parameter pH
pH resolution 0.001
mV resolution 0.1
Asymmetry 0 mV
Slope -59.2 mV