What to do if... Multi 3410
109
ba75863y18 01/2015
6 What to do if...
6.1 General information
Sensor symbol flashes
Display
Meter does not react to
keystroke
You want to know which
software version is in
the meter or IDS sensor
Cause Remedy
Calibration interval expired Recalibrate the measuring
system
Cause Remedy
Batteries almost empty Charge the batteries (see
section 3.3.2 C
ONNECTING
THE POWER PACK / CHARGING
THE BATTERIES)
Replace the batteries (see
section 5.1 M
AINTENANCE)
Cause Remedy
Operating condition undefined
or EMC load unallowed
Processor reset:
Press the <ENTER> and
<On/Off> key simultane-
ously
Cause Remedy
E. g., a question by the service
department
Switch on the meter.
Open the menu,
<ENTER
_> / Storage &
config / System / Service
information. The instrument
data are displayed.
or
Connect the sensor.
Press softkey [Info]/[More].
The sensor data are dis-
played (see section 4.5)