Multi 3410 Operation
78
ba75863y18 01/2015
4.9.4 Settings for IDS conductivity sensors
Overview The following settings are possible for IDS conductivity sensors:
Calibration record (display)
Calibration data storage (display)
Calibration interval
Reference temperature
Temperature compensation
TDS factor
Unit of the temperature
Settings The settings are made in the menu for the measured parameter, con-
ductivity. To open the settings, display the required parameter in the
measured value display and press the <MENU/ENTER> key. After
completing the settings, switch to the measured value display with
<M>.
The possible settings are individually displayed for each sensor. Below
the setting menu is displayed for two IDS sensors (TetraCon 925,
LR 925/01).
Default settings are printed in bold.
Setting menu of
TetraCon 925
Menu item Possible
setting
Description
Calibration /Cali-
bration record
- Displays the calibration record of
the last calibration
Calibration /Cali-
bration data stor-
age
- Displays the last calibration
records (max. 10)
Calibration /
Calibration data
storage / Output to
USB flash drive or
printer
- Outputs the calibration data stor-
age to a connected USB flash
drive/USB printer
Calibration /Cali-
bration interval
1 ... 150 ...
999 d
Calibration interval for the IDS
conductivity sensor (in days).
The meter reminds you to cali-
brate regularly by the flashing
sensor symbol in the measuring
window.