Operation Multi 3410
43
ba75863y18 01/2015
* Brand names or trade names are trademarks of their respective
owners protected by law.
Note
The buffers are selected in the menu, pH / <MENU/ENTER> / Calibra-
tion / Buffer (see section 4.6.3).
Calibration points Calibration can be performed using one to five buffer solutions in any
order (single-point to five-point calibration). The meter determines the
following values and calculates the calibration line as follows:
19
Precisa * 3.996
7.003
8.999
25 °C
20
Reagecon TEC * 2.000
4.010
7.000
10.000
25 °C
21
Reagecon 20 * 2.000
4.000
7.000
10.000
13.000
20 °C
22
Reagecon 25 * 2.000
4.000
7.000
10.000
13.000
25 °C
23 Chemsolute * 2.000
4.000
7.000
10.000
20 °C
24
USABlueBook
* 4.000
7.000
10.000
25 °C
25 YSI * 4.000
7.000
10.000
25 °C
No. Buffer set* pH values at