Operation Multi 3410
47
ba75863y18 01/2015
Note
For single-point calibration, the instrument uses the Nernst slope
3 Thoroughly rinse the IDS sensor with deionized water.
4 Immerse the IDS pH sensor in buffer solution 1.
5 For measurements without temperature sensor
(e.g. when using an IDS adapter
):
Measure the temperature of the buffer manually and enter it with
<><>.
6 Start the measurement with <MENU/ENTER>.
The measured value is checked for stability (stability control).
The [AR] status indicator is displayed. The measured parameter
flashes.
7 Wait for the end of the measurement with stability control or
accept the calibration value with <MENU/ENTER>.
The calibration display for the next buffer appears (voltage dis-
play).
8 If necessary, finish the calibration procedure as a single-point cal-
ibration with <M>.
The calibration record is displayed.