EasyManuals Logo

Xylem wtw Multi 3410 User Manual

Xylem wtw Multi 3410
126 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #76 background imageLoading...
Page #76 background image
Multi 3410 Operation
76
ba75863y18 01/2015
Note
You can prematurely terminate the Stability control function manually
with <MENU/ENTER> at any time. If the Stability control function is pre-
maturely terminated, the current measurement data are output to the
interface without the AutoRead info.
Criteria for a stable
measured value
The Stability control function checks whether the measured values are
stable within the monitored time interval.
The minimum duration until a measured value is assessed as stable is
the monitored time interval. The actual duration is mostly longer.
3 Using <MENU/ENTER>, start a further measurement with Sta-
bility control.
or
Release the frozen measured value again with <AR>.
The display switches to the measured value display.
The [AR][HOLD] status display disappears.
Measured
parameter
Time interval Stability in the time interval
Conductivity ϰ 10 seconds
Δ
: Better than
1.0% of measured value
Temperature 10 seconds
Δ
: Better than 0.5 °C

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Xylem wtw Multi 3410 and is the answer not in the manual?

Xylem wtw Multi 3410 Specifications

General IconGeneral
BrandXylem
Modelwtw Multi 3410
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals