Multi 3410 Operation
80
ba75863y18 01/2015
Setting menu of
LR 925/01
Menu item Possible set-
ting
Description
Cell constant 0.090 ... ...
0.100 ...
0.110 cm
-1
The cell constant can be dis-
played and set here
Temp. comp. (TC) /
Method
nLF
Lin
off
Procedure for temperature
compensation (see section
4.9.3).
This setting is only available
for the measured parame-
ters, ϰ and
ρ.
Temp. comp. (TC) /
Linear coeff.
0.000 ...
2.000 ...
10.000 %/K
Coefficient of the linear tem-
perature compensation.
This setting is only available
when the linear temperature
compensation is set.
Temp. comp. (TC) /
Reference temp.
20 °C
25 °C
Reference temperature
This setting is only available
for the measured parame-
ters, ϰ and
ρ.
TDS factor 0.40 ... 1.00 Factor for TDS value
Reset - Resets all sensor settings to
the delivery condition (see
section 4.14.1)