Multi 3410 Operation
82
ba75863y18 01/2015
Determining the cell
constant
For this calibration procedure, the Type setting must be set to cal . Pro-
ceed as follows to determine the cell constant:
Error Error
Outside the range
0.450 ... 0.500 cm
-1
Eliminate the error according to
chapter 6 W
HAT TO DO IF...
Display Calibration
record
Cell constant [cm
-1
]
1 Connect an IDS conductivity sensor to the meter.
2 In the measured value display, select the conductivity parame-
ter with <M>.
3 Start the calibration with <CAL>.
The cell constant that was calibrated last is displayed.
4 Immerse the IDS conductivity sensor in the control standard
solution, 0.01 mol/l KCI.
5 Start the measurement with <MENU/ENTER>.
The measured value is checked for stability (stability control).
The [AR] status indicator is displayed. The measured parame-
ter flashes.
6 Wait for the measurement with stability control to be completed
([HOLD][AR]) display or
accept the calibration value with <MENU/ENTER>.
The calibration record is displayed and output to the interface.
7 Switch to the measured value display with <MENU/ENTER>.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Xylem wtw Multi 3410 and is the answer not in the manual?

Xylem wtw Multi 3410 Specifications

General IconGeneral
BrandXylem
Modelwtw Multi 3410
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals