EasyManua.ls Logo

Xylem YSI 2500 - Page 47

Xylem YSI 2500
98 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
6-24
necessary, touch the [Flush] button to flush the sample module again. Note that when the instrument is first powered up, it
may take several hours for the baseline currents to drop below 6 nA.
6.2.2.3 Calibrate
The 2500 Analyzer will automatically calibrate before running a sample batch. You may initiate manual calibration from
the [Modules] tab of the Service screen by touching the [Calibrate] button.
Calibration status is displayed on the screen.
The instrument attempts to calibrate each active probe up to 5 times before aborting calibration. If calibration fails, see
Section 10 Troubleshooting.

Table of Contents

Related product manuals