EasyManua.ls Logo

Xylem YSI 2500 - Module

Xylem YSI 2500
98 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
6-23
Touch [Save] to confirm the position.
6.2.2 Module
From the Service menu, touch the [Module] tab.
The Module tab displays the status of the probes, probe current for each enzyme probe, and the temperature. The probe
current is expressed in nA (nanoamperes, 10
-9
amperes), a very low level of electrical current.
6.2.2.1 Sipper Pump
Home
Touch the [Home] button to Home the sipper pump plunger. The pump plunger will extend fully, then retract slightly.
Aspirate
Touch the [Aspirate] button. The pump plunger will retract about half way as it does when it aspirates a sample. Note that
the actual distance depends on the sample size setting.
Dispense
Touch the [Dispense] button. The pump plunger will extend about half way as it does when it dispenses a sample into the
sample module.
When you have finished testing the sipper pump, touch the [X] button to return to the main display.
6.2.2.2 Flush
Touch the [Flush] button to flush the sample module with buffer. Observe the probe current values (baseline). If they are
above 6 nA, check to see if they are decreasing in value. Check the sample module; it should be full of buffer. If

Table of Contents

Related product manuals