1-1
1 Introduction
Description 1.1
The 2500 Biochemistry Analyzer is a laboratory instrument intended for use in research, food-processing and
bioprocessing applications. THE 2500 Analyzer IS NOT FOR HUMAN MEDICAL DIAGNOSTIC USE OR FOR HUMAN
PERFORMANCE EVALUATION.
The 2500 Analyzer can be set up to measure glucose and/or lactate in a sample.
User Features
Easily expand analytes or chemistries
Multiple units use much less bench space
Proprietary enzyme electrode
Fast, accurate, and analyte-specific results
Uses biological separation technology
No hazardous chromatography solvents to dispose of
Icon-driven user interface with touchscreen
Save data on a USB drive, send it over the network,
or access it in a searchable database anytime
Minimizes operator learning curve
General Specifications 1.2
Sample size ................................................. Adjustable from 10 to 50 microliters (aspirated volume), per module
Response time ............................................ Enzyme Sensors
• Sample results in 60 seconds (average)
• Complete sample-to-sample cycle in less than 2 minutes
(May vary with analyte and sample matrix.)
Output signals:
Serial .......................................................... USB and RS232
Power requirement ............................... 100–240 VAC ±10%
50–60 Hz ±5%
42 Watts nominal
Working environment:
Ambient temperature ................................ 15–35°C
Relative humidity ...................................... 10–75% (non-condensing)
Regulatory compliance .................................. ETL, CE, RoHS
61010-1 compliance:
• Pollution degree 2
• Installation category 2
• Altitude 2000m
• Atmosphere 75 KPa to 106 KPa
• Indoor use only
Instrument dimensions ................................ 12" wide x 20.5" deep x 15.75" high (30.5 cm x 52.1 cm x 40 cm)