15-1
15 Appendix B – Concentration Unit Conversion
In the 2500 Analyzer Batch menu, you have the option to assign units of concentration. There are default values set
based on calibration solutions offered by YSI. Below is a table of unit conversions for these calibration solutions.
If you are using a standard of a value not listed in the preceding table, refer to the example conversions below to help
calculate your unit of choice.
Example Conversions
Beginning with 2.50 g/L glucose, convert this to mg\L, then % and finally to mmol/L:
1. Multiply by unit conversion(s)
2. Cancel units common in “numerator” and “denominator”
3. Multiply numbers
? mg/L = 2.50 g/L
= (2.50 g/L)(1000 mg/g) = (2.50)(1000) mg/L = 2500 mg/L
? % (w/v) = 2.50 g/L
= (2.50 g/L)(0.1 L/dL) = (2.50)(0.1) g/dL = 0.250 g/dL
= 0.250 % (Note: g/dL is g/100ml or percent)
? mmol/L = 2.50 g/L
= (2.50 g/L)(1 mole/180 g)(1000 mmole/mole)
= (2.50)(1/180)(1000) mmole/L = 13.89 mmol/L
Linearity Test. Concentration Unit Conversion 15.1
NOTE: The linearity concentration ranges for each chemistry are shown (top to bottom) as upper limit, theoretical, and
lower limit for each of five concentration units.
FCN Membrane Integrity Test. Concentration Unit Conversion 15.2
NOTE: Use the values from the preceding tables only when calibrating with the appropriate YSI calibrator solution
Glucose (2776 or 2747) and L-Lactate (2776).