15-4
IM 2558A-01EN
Communication Execution Errors (200 to 299)
Code Message Corrective Action Page
203 Command protected See “Adjustment and Calibration“ in section 15.8. 15-11
221 Setting conflict. Check settings that are related to each other. Chapter 13
222 Data out of range. Check the ranges of the settings. Chapter 13
223 Too much data. Check data byte lengths. Chapter 13
224 Illegal parameter value. Check the ranges of the settings. Chapter 13
225 OverFlow. Keep program messages to 1024 bytes or less in length, including
<PMT>.
12-2
226 Out Of Memory. Keep program messages to 1024 bytes or less in length, including
<PMT>.
12-2
241 Hardware missing. Check that the specified options are all installed. —
260 Expression error. Mathematical operations cannot be used. —
270 Macro error. The 2558A does not support the IEEE 488.2 macro specifications. —
272 Macro execution error. The 2558A does not support the IEEE 488.2 macro specifications. —
273 Illegal macro label. The 2558A does not support the IEEE 488.2 macro specifications. —
275 Macro definition too long. The 2558A does not support the IEEE 488.2 macro specifications. —
276 Macro recursion error. The 2558A does not support the IEEE 488.2 macro specifications. —
277 Macro redefinition not allowed. The 2558A does not support the IEEE 488.2 macro specifications. —
278 Macro header not found. The 2558A does not support the IEEE 488.2 macro specifications. —
15.3 Communication Error Messages