<4. Setting converter>
4-10
IM 12A01F01-03EN 2nd Edition : Feb. 19, 2019-00
l Process parameter

appropriately, a setting error occurs.
Selecting appropriate process parameter according to sensors to use.
Common to all: temperature
for PH: pH, ORP, rH
for SC Conductivity-TC1, Conductivity-TC2, Resistivity-TC1, Resistivity-TC2
Concentration TC1, Concentration TC2
for DO Oxygen
for Converter: AI, script *1
󰀨󰀨󰀨󰀨
󰀨
Average pH, Average ORP, Average CONDUCT, Average RESIST,
Average Oxygen
Ratio, Passage (%), Rejection (%), Deviation (%),

*1: Latest revision does not support script. Do not select script.
*2: When you connect multiple sensors, you will gain the arithmetic from Sensor 1-1 and Sensor 2-1. In order to gain this value, set
same type of sensors to Sensor 1-1 and Sensor 2-1, otherwise a setting error will be detected. Note that latest revision does not
support the arithmetic out of multi sensors.
l output setup
Select one of the output methods, Linear or Table.
Linear: Set 0% value and 100% values.
 
l Sub linear 0%, Sublinear 100%
Set 0% and 100% value at Contact input change range.
l Burn
󰀨
failure. Read 4.7 Error settings for setting failures.
󰀨 
 
 
Burn happens only when failure is detected on the sensors you set on Process parameter slot.
Sensors you did not set on Process parameter slot do not output burn even when their failure is
detected.
Whenever the converter’s failure is detected, it outputs burn value.
If a communication error inside of the converter is detected, the converter seeks the error

error, burn is output.
l Damping time

(attenuation time). Set this time in sec.